Unit passed displacement test.
Sleep current measurement and active current measurement within specifications.
The power management tool indicated abnormal behavior of the lithium backup battery loaded voltage (loaded vlith) as shown on the power management graph and confirmed power error, low battery alarm due to connector resistance issue.
After disconnecting and reconnecting the internal battery connector, the insulin pump was monitored and functioned properly.
Unit received with cracked retainer, minor scratched display window, fading serial number label and scratched case.
|