Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/18/2018
Event Type  Malfunction  
Manufacturer Narrative

A steris service technician arrived onsite to inspect the sterilizer and found that the heat exchanger required replacement and the drain piping connected to the unit required repair. The technician replaced the heat exchanger, repaired the drain piping, ran a test cycle, and confirmed the unit to be operating according to specifications. The unit was manufactured in 2013 and is under steris service agreement for maintenance activities. No additional issues have been noted.

 
Event Description

The user facility reported that the amsco 400 sterilizer was leaking water onto the floor. No report of injury, procedure delay, or cancellation.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAMSCO 400 STERILIZER
Type of DeviceSTERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX 67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX 67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key7700074
MDR Text Key114695297
Report Number3005899764-2018-00063
Device Sequence Number1
Product Code FLE
Combination Product (Y/N)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type USER FACILITY
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Type of Report Initial
Report Date 07/18/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received07/18/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/18/2018
Was Device Evaluated By Manufacturer? Yes
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-