• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 06/27/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.No parts have been returned to the manufacturer for analysis.Parts not returned.
 
Event Description
Medtronic received information that when in use, the navigation system's surgeon monitor appeared to flicker and loose signal.The rep was able to temporarily get the monitor to function by disconnecting and re-connecting the main lemo video/power cable to the monitor but after a few seconds the monitor would stop working.There was no patient present when this issue was identified.
 
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that issue was unable to be replicated.The system then passed the system checkout and was found to be fully functional.
 
Manufacturer Narrative
Correction: adding proper coding.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
koko tyler
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208903200
MDR Report Key7700275
MDR Text Key114413753
Report Number1723170-2018-03563
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 08/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received08/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/04/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-