Model Number 2090W
Device Problems Loss of Power (1475); Unexpected Shutdown (4019)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
Event Date 07/03/2018
Event Type  malfunction  
Manufacturer Narrative
Product analysis: manufacturer's analysis was unable to confirm the customer comment that the programmer crashed.The programmer passed incoming functional testing.However, errors were found in the programmer logs.It was also indicated that the keyboard and a latch were broken.The programmer was repaired and returned to service.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer would crash when attempting to check device.The programmer was returned to service.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7704277
MDR Text Key114802066
Report Number2182208-2018-01322
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00613994127631
UDI-Public00613994127631
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090W
Device Catalogue Number2090W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/14/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-