MAUDE Adverse Event Report: MEDTRONIC MEXICO PACIFIC XTREME; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO PACIFIC XTREME; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Catalog Number PCF030120130
Device Problems Burst Container or Vessel (1074); Insufficient Information (3190)
Patient Problems No Consequences Or Impact To Patient (2199); Rupture (2208)
Event Date 06/20/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
The physician was attempting to use a pacific xtreme balloon to treat a moderately calcified plaque lesion (60% stenosis) in the superficial femoral artery.The artery had no tortuosity.The device was prepped with no issues identified.Spider fx embolic protection was used.The device did not pass through a previously deployed stent and no excessive force was used.No resistance was encountered.During the first balloon inflation with a pressure pump at 7 atm a circumferential/radial burst occurred.The procedure was completed using another balloon the same make and size.
 
Manufacturer Narrative
Additional information: the damage in the equipment was observed during the preparation process and after opening the device.The top could not pass through the guide wire.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACIFIC XTREME
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
MEDTRONIC MEXICO
avenida paseo del cucapah #105
parque industrial el lago
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
avenida paseo del cucapah #105
parque industrial el lago
tijuana,bc 22570
MX   22570
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key7705619
MDR Text Key114812630
Report Number9612164-2018-01803
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K103464
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/13/2020
Device Catalogue NumberPCF030120130
Device Lot Number214581210
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2018
Date Device Manufactured12/13/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age64 YR
-
-