Model Number MMT-751NAS
Device Problems Improper or Incorrect Procedure or Method (2017); Device Displays Incorrect Message (2591)
Patient Problem Hypoglycemia (1912)
Event Date 05/03/2018
Event Type  malfunction  
Manufacturer Narrative
Pump passed the rewind, basic occlusion, prime/a33 and displacement test. Pump was received with stuck motor error alarm loop during bolus delivery due to motor encoder signal out of phase. Unable to perform the occlusion and no delivery test due to motor error alarm. Unable to confirm e70 alarm due to erased history file.
 
Event Description
The customer reported via phone call the insulin pump alarmed motor error after exposure to mri and had a low blood glucose event. The customer¿s blood glucose level was 52 mg/dl at the time of the incident. The customer stated that the drive support cap was normal and that the insulin pump was exposed to high magnetic fields. Customer reported a motor position encoder error. Customer had a low blood glucose of 52 mg/dl and treated with food. No low blood glucose troubleshooting was performed. The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan. The insulin pump is being replaced and is expected to return for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name530G INSULIN PUMP MMT-751NAS
Type of DeviceARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key7712662
MDR Text Key114854159
Report Number3004209178-2018-94971
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation
Type of Report Initial
Report Date 07/23/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberMMT-751NAS
Device Catalogue NumberMMT-751NAS
Device Lot NumberA4751NASJ
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/03/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-