• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616
Device Problems Power Cord (497); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Malfunction  
Manufacturer Narrative

There was a relationship found between the returned device and the reported incident. A visual inspection was performed on the exterior of product and no physical damage was observed. Mdu failed functional tests with high current readings and overheating when power cord was bent or twisted. Power cord assembly grew warm during testing. After troubleshooting, the cause of the overheating was observed to be a defective power cord assembly. A visual inspection was performed on the power cords external covering and no physical damage was observed. It was determined that the power cord has a shorted or open internal wiring. Motor and hall board were tested and passed functional testing. The complaint investigation has concluded that this unit has succumbed to physical damage to power cord. Factors which can contribute to a shorted power cord include rolling a heavy cart over the cord or excessive bending of the cord. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.

 
Event Description

It was reported that mdu is not working. Product evaluation determined power cord gets hot. No delay or patient injury was reported.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key7715207
MDR Text Key115194885
Report Number1643264-2018-00553
Device Sequence Number1
Product Code HAB
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE
Reporter Occupation
Type of Report Initial
Report Date 07/24/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received07/24/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue Number72200616
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer06/21/2018
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/16/2018
Was Device Evaluated By Manufacturer? Yes
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-