COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 383400
Device Problems No Display/Image (1183); Image Display Error/Artifact (1304)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 05/29/2018
Event Type  malfunction  
Manufacturer Narrative
An evaluation of the kangaroo pump was performed.The unit was triaged and the reported issue could not be confirmed at this time; the display lcd was legible but had water mark in the background all device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that an issue occurred with an enteral feeding pump.Upon receiving an email on (b)(6) 2018 the customer reported the screen was not legible.There was no patient harm or medical intervention.
 
Manufacturer Narrative
The unit was triaged and the customer¿s reported condition was confirmed.The pump was excessively damaged, so the root cause is categorized as customer misuse.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
building 10- no 789 puxing roa
shanghai 20111 4
CN  201114
MDR Report Key7718334
MDR Text Key115280559
Report Number3006451981-2018-00521
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 10/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date01/14/2016
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2018
Date Manufacturer Received08/08/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-