• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 382400
Device Problems No Audible Alarm (1019); Appropriate Term/Code Not Available (3191)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, the feeding pump has no volume.No further information was available.The event was discovered during their testing, and there is no reported patient involvement.
 
Manufacturer Narrative
An evaluation of the kangaroo epump was performed for the reported condition of, ¿the feeding pump has no volume¿.The unit was triaged and the reported condition was confirmed.Internal inspection and investigation reveals a damaged component u14 on the pcba.A trend has been identified and a correction action has been opened to address this issue.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
building 10- no 789 puxing roa
shanghai 20111 4
CN  201114
MDR Report Key7720643
MDR Text Key115272773
Report Number3006451981-2018-00524
Device Sequence Number1
Product Code LZH
UDI-Device Identifier10884521006614
UDI-Public10884521006614
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 10/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date12/21/2014
Device Model Number382400
Device Catalogue Number382400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2018
Date Manufacturer Received08/08/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-