Model Number 382400 |
Device Problems
No Audible Alarm (1019); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, the feeding pump has no volume.No further information was available.The event was discovered during their testing, and there is no reported patient involvement.
|
|
Manufacturer Narrative
|
An evaluation of the kangaroo epump was performed for the reported condition of, ¿the feeding pump has no volume¿.The unit was triaged and the reported condition was confirmed.Internal inspection and investigation reveals a damaged component u14 on the pcba.A trend has been identified and a correction action has been opened to address this issue.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|