The returned device, intended for use in treatment, was returned as an mdr for evaluation.There was a relationship found between the device and the reported incident.Visual inspection shows a device which was returned with an open bracket.The self-capture was received seperately.There were no manufacturing abnormalities found on the device.During functional evaluation the two step trigger performed as intended.The needle deploy through a om-3003 dome foam modelby using the index finger and depressing the trigger, however the self-capture part is missing.The complaint was verified and the root cause could not be determined with certainty.Excessive force should not be applied to the device when manipulating soft tissue, bone, or hard objects.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
|