Model Number FUSION
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 07/05/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Missing udi number.Missing manufacture date.Parts not returned.
 
Event Description
Medtronic received information regarding a navigation device being used outside a procedure.It was reported that the site was unable to boot the system into the application.The system would display the first medtronic screen and then go black.Menu bar at bottom of screen could not be pulled up.The site attempted multiple boot ups with the axiem connected and disconnected with no resolution.The device was able to enter the bios menu, but when the site exited it still would not completely boot.The initial medtronic screen would flash for a second and then the screen would go black and become unresponsive.There was no patient present when this issue was identified.
 
Manufacturer Narrative
Additional information: udi and manufacture date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
jack edell
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7722905
MDR Text Key115191108
Report Number1723170-2018-03724
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/05/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-