MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ANTHEM RF CRT-P IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ANTHEM RF CRT-P IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3210
Device Problems Pacing Inadequately (1442); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/02/2018
Event Type  malfunction  
Manufacturer Narrative
Correction: the information previously submitted - initial reporter was incorrect. The information provided in this supplemental report is correct and should supersede the previously reported information.
 
Event Description
New information received indicated that the patient underwent an unrelated procedure during the episode. During the procedure, the pacemaker was temporarily reprogrammed to a low base rate which resulted in the long pacing delay episode. No further device intervention was performed.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis. Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented remotely via merlin. Net. Review of transmissions revealed that the pacemaker exhibited long pacing delays due to periodic switches in pacing rate. It was noted that the patient had their base rate programmed low during the episode. Patient was asymptomatic. No further information was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameANTHEM RF CRT-P
Type of DeviceIMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7724890
MDR Text Key115613527
Report Number2017865-2018-11007
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 07/31/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date03/31/2015
Device Model NumberPM3210
Device Lot Number4259055
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/04/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/22/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-