• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Device Problems Pumping Stopped (1503); Mechanical Jam (2983)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/04/2017
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp) via a clinical study regarding a patient receiving fentanyl 100mcg/ml for a total dose of 35. 98mcg/day, bupivacaine 30mg/ml for a total dose of 10. 794mg/day, and clonidine 600mcg/ml for a total dose of 215. 89mcg/day via an implantable pump for non-malignant pain. It was reported on (b)(6) 2017 a review of pump logs/device interrogation revealed multiple motor stalls and recoveries had occurred. It was noted that a pump motor stall occurred on (b)(6) 2017 at 23:27 with a motor stall recovery on (b)(6) 2017 at 00:04 and a pump motor stall occurred on (b)(6) 2017 at 14:40 with a motor stall recovery at 15:01. There was no report of an magnetic resonance imaging (mri). No actions taken. The outcome of the event resolved without sequelae on (b)(6) 2017. The device diagnosis was pump motor stall. The etiology of the event indicated the relationship of the event to the device or therapy was related and indicated the relationship of the event to the implant procedure was not related. The event date was (b)(6) 2017. No further complications were reported/anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from an hcp via a clinical study on (b)(6) 2018. It was reported that the cause of the stall was not determined.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7725218
MDR Text Key115801804
Report Number3004209178-2018-16792
Device Sequence Number0
Product Code LKK
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,study
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received07/26/2018
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/28/2017
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured01/13/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-