Model Number 10-80-00
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/02/2018
Event Type  Malfunction  
Manufacturer Narrative

Livanova (b)(4) manufactures the s5 roller pump. The incident occurred in (b)(6). A livanova field service representative was dispatched to the facility to investigate. The service representative was able to confirm the reported failure and replaced the touch screen to resolve the malfunction. Subsequent functional verification testing was completed without further issues and the unit was returned to service. Through follow-up communication, livanova (b)(4) learned that liquid had entered the touchscreen. A photo was provided which confirmed this issue. A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue. This is a known issue and corrective action has been implemented.

 
Event Description

Livanova (b)(4) received a report that the touch screen of an s5 roller pump became unresponsive during a procedure. There was no report of patient injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameS5 ROLLER PUMP
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7729097
MDR Text Key115366789
Report Number9611109-2018-01117
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeIT
PMA/PMN NumberK071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Remedial Action Other
Type of Report Initial
Report Date 07/27/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received07/27/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number10-80-00
Was Device Available For Evaluation? No Answer Provided
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/02/2018
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured11/16/2009
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-