Catalog Number IGTCFS-65-1-FEM-CELECT-PT
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/05/2018
Event Type  Malfunction  
Manufacturer Narrative

(b)(4). Name and address for importer site: (b)(4). Type of investigation not yet determined. Investigation is still in progress.

 
Event Description

Description of event according to initial reporter: "physician contacted me yesterday during filter placement procedure. Was placing from left femoral vein and encountered resistance while trying to advance filter. Tried to pull back, which caused sheath to split. Removed the whole system and tried again with a new femoral celect platinum set. Same thing happened, so he removed the whole system again. This time he put in another manufacturers filter which traversed fine. He said that the other manufactures sheath was thicker, so maybe that is why filter advanced through with no issues?" patient outcome: patient was fine. No additional procedure needed.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCOOK CELECT® PLATINUM NAVALIGN FEMORAL VENA CAVA FILTER SET
Type of DeviceDTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA 4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov DK-46 32
DA DK-4632
Manufacturer Contact
thomas hessner kirk
sandet 6
bjaeverskov DK-46-32
DA   DK-4632
56868686
MDR Report Key7730771
MDR Text Key115574702
Report Number3002808486-2018-00831
Device Sequence Number1
Product Code DTK
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK121629
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL,USER FACI
Reporter Occupation
Type of Report Initial,Followup
Report Date 10/23/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received07/30/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue NumberIGTCFS-65-1-FEM-CELECT-PT
Device LOT NumberUNKNOWN
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Distributor Facility Aware Date07/06/2018
Event Location No Information
Date Manufacturer Received10/01/2018
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-