Batch review performed on 26 july 2018: lot 083125: (b)(4) items manufactured and released on 11 december 2008.Expiration date: 2013-10-31.No anomalies found related to the issue.To date, (b)(4) items of the same lot have been already sold without any similar reported event.On 27 july the medical affairs director made the following analysis: 9 years after primary cementless double mobility tha the mobile liner was revised to a new one, possible because of wear.The age and activity level of the patient are not specified.Such a duration is compatible with literature reports for standard pe mobile liners.The osteolytic process does not seem to have compromise fixation of metal components, which, according to report, have not been revised.We tend to categorize this event as regular wear and not as an accident.
|