• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Electrical Power Problem (2925)
Patient Problem No Patient Involvement (2645)
Event Date 06/20/2018
Event Type  Malfunction  
Manufacturer Narrative

The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601). Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). A livanova field service representative was dispatched to the facility to investigate. The service representative was able to confirm the reported issue and traced the failure to a faulty compressor. The repair was determined to be non-economical and no further service action was performed. A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report. Device not returned.

 
Event Description

Livanova (b)(4) received a report that a heater-cooler system 3t triggered a circuit breaker during a disinfection procedure. There was no patient involvement.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7737624
MDR Text Key115649665
Report Number9611109-2018-01124
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Remedial Action Other
Type of Report Initial
Report Date 07/31/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received07/31/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number16-02-80
Was Device Available For Evaluation? No Answer Provided
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received06/20/2018
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured01/26/2015
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-