• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/09/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site at the time of filing.Onsite functional and visual examination was performed by a manufacturer representative.The issue was resolved after parts were replaced.The computer was returned to the manufacturer for analysis.Functional testing found that upon reboots the computer started to power cycle.The computer was found to be defective.
 
Event Description
Medtronic received information regarding a navigation device being used for a sacroiliac and thoracolumbar procedure.It was reported that the system became unresponsive temporarily.The surgeon went to navigate and unresponsiveness occurred again.Rebooting resolved the reported issue.Site reported this issue has happened various times in the past.There was no impact to patient.There was a reported delay to the procedure of 5 minutes due to this issue.
 
Manufacturer Narrative
Patient information received.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Patient information received.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
parker desautel
navigation customer quality
826 coal creek circle
louisville, CO 80027
MDR Report Key7738440
MDR Text Key115659067
Report Number1723170-2018-03768
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberS7
Device Catalogue Number9733927
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/05/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age37 YR
Patient Weight79
-
-