• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HOL ML 5 MILLIMETER ENDO APPLIER; APPLIER, SURGICAL, CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HOL ML 5 MILLIMETER ENDO APPLIER; APPLIER, SURGICAL, CLIP Back to Search Results
Catalog Number 544965
Device Problems Separation Failure (2547); Difficult to Open or Close (2921)
Patient Problem No Information (3190)
Event Date 07/09/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that the clip applier was used during laparoscopy cholecystectomy.After locking the clip around the vessel, the jaws of the applier remained closed and would not release.The patient's condition is unknown at this time.
 
Manufacturer Narrative
(b)(4).The dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was manufactured at the (b)(4) as part of a 50 pc.Lot in (b)(6) of 2017.The returned instrument was evaluated as received and found that the handle to jaw mechanism was sluggish/binding and knob rotation mechanism was dry and sluggish feeling and the jaws would remain closed in multiple locations when closed thus we are able to validate this complaint.After the initial evaluation this instrument was properly lubricated and further evaluation showed that this instrument picks up, retains, closes and releases multiple clips as required of its function both with and without the use of silastic test tubing.Parts were 100% visually inspected and function tested at the (b)(4) before instruments were sent to customer.No irregularities were found and or reported at the time of inspection and assembly of the product, as this is a standardized process for all instruments manufactured at this facility.At this time it is un-determined what caused the alleged issue, but it is suspected that the customer did not "lubricate" the instrument prior to sterilization as recommended in ifu (l06109 r04) supplied with the instrument which states "the instrument must be cleaned, lubricated, functionally checked, and sterilized prior to each use.Use a non-silicone, water-based lubricant prior to sterilization." no corrective action required at this time.
 
Event Description
It was reported that the clip applier was used during laparoscopy cholecystectomy.After locking the clip around the vessel, the jaws of the applier remained closed and would not release.The patient's condition is unknown at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOL ML 5 MILLIMETER ENDO APPLIER
Type of Device
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
3015 carrington mill blvd
morrisville NC 27560
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key7738603
MDR Text Key115668396
Report Number3011137372-2018-00222
Device Sequence Number1
Product Code GDO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number544965
Device Lot Number06H1638572
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-