Getinge usa sales, llc (importer) is submitting this report on behalf of maquet critical care ab (manufacturer).Ref.Exemption #: (b)(4).Getinge usa sales, llc (b)(4).Contact peron: (b)(6).Our field service engineer confirmed the fault during on-site investigation.The air gas module was found faulty and was replaced.The returned air gas module was simulated use tested in a reference anesthesia system.The system checkout performed initially failed on the gas supply test and the flow transducer test.The tests failed due to that the air gas module measured a lower air supply pressure than the actual wall supply pressure was.The allowed air supply pressure range is 2.5-6.5 bar.During test in ventilation mode, alarms for air supply pressure low were generated on and off.The ventilation mode test was ended after 2 days.Several system checkouts were performed after this and all tests passed without deviations.Evaluation of the received device logs confirm the reported alarms for air supply pressure low and the failing tests during system checkout.Our conclusion is that the air gas module was the cause of the reported failure.The root cause of the air gas module failure has not been determined since the failure is intermittent.
|