Model Number 9438-06
Device Problem Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/26/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018, that on (b)(6) 2018, a pairing failure between the transmitter and smart device and/or receiver occurred.No additional patient or event information is available.Data was provided for evaluation.Data review confirmed the reported event of pairing failure.The root cause was determined to be a failed transmitter error.Reported issue of pairing failure is reportable based on the finding of transmitter failure error.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The complaint device was returned for evaluation.An exterior physical visual inspection was performed and the device passed.The transmitter voltage was tested and failed at 0vdc.The cloud/share data was available.Transmitter fatal error was found in share log but not related to complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key7741635
MDR Text Key115788386
Report Number3004753838-2018-078471
Device Sequence Number1
Product Code PFQ
UDI-Device Identifier10386270000221
UDI-Public10386270000221
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/18/2018
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Date FDA Received08/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number9438-06
Device Catalogue NumberSTT-GF-001
Device Lot Number6006057
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/23/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured12/20/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age33 YR
-
-