MAUDE Adverse Event Report: MAQUET CV HS III PROXIMAL SEAL SYTEM 3.8MM CLAMP, VASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CV HS III PROXIMAL SEAL SYTEM 3.8MM CLAMP, VASCULAR Back to Search Results
Catalog Number C-HSK-3038
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/28/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4). A lot history record review was completed for the reported product lot number. There were no ncmr¿s for the reported lot number. The device has been returned to the factory and is being evaluated. A supplemental report will be submitted when the evaluation is completed.
 
Event Description
The hospital reported that during a coronary artery bypass procedure, hs iii proximal seal system 3. 8mm failed to deploy by its hardness. A replacement device was used to complete the procedure. The hospital did not report any patient effects.
 
Manufacturer Narrative
(b)(4). The device was returned to the factory for evaluation. A visual inspection and a photographic inspection was conducted. Signs of clinical use and evidence of blood was observed on the delivery device. The loading device was not returned for evaluation. The blue slide lock was disengaged and the white plunger was fully depressed on the delivery device. The seal and tension spring assembly were returned outside the delivery device. The seal was observed to be unraveled with no defects. The tether is observed to be cut in an angle. Blood was observed inside the delivery device, as well as on the seal, indicating that there was an attempt to insert the device into the aorta. No measurements could be taken of the delivery tube due to the presence of blood. Based on the return condition of the device, the reported failure ¿positioning issue¿ was not confirmed.
 
Event Description
The hospital reported that during a coronary artery bypass procedure, hs iii proximal seal system 3. 8mm failed to deploy by its hardness. A replacement device was used to complete the procedure. The hospital did not report any patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHS III PROXIMAL SEAL SYTEM 3.8MM
Type of DeviceCLAMP, VASCULAR
Manufacturer (Section D)
MAQUET CV
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
MAQUET CV
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
45 barbour pond drive
wayne, NJ 07470
MDR Report Key7742300
MDR Text Key116069016
Report Number2242352-2018-00734
Device Sequence Number1
Product Code DXC
Combination Product (y/n)N
PMA/PMN Number
K080169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 08/01/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date02/01/2019
Device Catalogue NumberC-HSK-3038
Device Lot Number25136926
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Device Age YR
Event Location No Information
Date Manufacturer Received09/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-