An investigation was performed in response to a backup failure observed on one of the vitek 2 systems pcs at the customer site, and was followed by an internal system error (ise).Local customer service (lcs) recommended restoring data backups from a previous day.Results from cards originally tested on (b)(6) 2018 were not available, and testing had to be repeated.A review of the logs provided from the dates of occurrences of (b)(6) 2018 indicate that the cause of the problem is an invalid certificate setup causing communication failures to the biomérieux user management service for the purposes of card processing, endofday and manual backups and restores.The impact of not having the system in a proper controlled/validated state can be as simple as minor errors, to inability to login, to loss of card data - depending on the usage of the system in the improper state.It is recommended that sites be properly setup by running the certificate setup scripts for the applicable setup type as indicated in the system setup guides.And that sites be notified that modifying the system in ways such as changing the pc name or modifying the network configuration can cause the certificates to become invalid.
|