Boston scientific received information that the patient with this cardiac resynchronization therapy pacemaker (crt-p) presented to the emergency room with dizzy spells.
The device and a non-boston scientific right ventricular (rv) lead exhibited noise, oversensing, pacing inhibition of approximately two seconds, and impedance measurements of greater than 2000 ohms.
It was questioned if the noise could have been due to minute ventilation (mv) oversensing, although it was noted that the noise did not look like mv signals.
This feature was programmed off.
Patient isometrics were performed which did not reproduce any noise.
The patient had also experienced a syncopal event.
There were no events stored in the device at the time of the syncope; however, it was noted that the events may have been overwritten.
Programming options were discussed at that time.
A lead fracture was suspected.
A surgical revision was performed and the non-boston scientific rv lead was surgically abandoned and replaced.
This device remains in service with the new lead.
No additional adverse patient effects were reported.
|