Model Number 1420
Device Problems No Display/Image (1183); Loss of Power (1475); Power Problem (3010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/09/2018
Event Type  malfunction  
Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Other devices involved in this event: heartware ventricular assist system ¿ battery / (b)(4) / model #: 1650 / expiration date: 2016-10-31 udi #: (b)(4) mfg date: 2015-10-31 (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the ventricular assist device (vad) controller exhibited a loss of power and the screen went blank.It was further reported that power switching also occurred with the battery associated with the loss of power to the controller.Lubrication servicing was performed on the battery.The controller and battery remain in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product event summary:the controller and battery were not returned for evaluation.The log files of the controller in use during the reported event revealed that the controller contained a feature that records whether a power source experienced a communication error or a disconnection within each 15-minute interval.Review of the controller event file revealed a controller power-up on (b)(6) 2018 at 17:32:16 followed by motor start at 17:32:20.The loss of power was for 13 seconds.The power sources attached before the loss of power, which was recorded at 17:17:27, were battery on power port 1 with 22% relative state of charge (rsoc) and a battery on power port 2 with 19% rsoc.The power sources attached after the loss of power, which was recorded at 17:32:52, were battery on power port 1 with 92% rsoc and no power source on power port 2.Log file analysis also revealed 2 power switching events due to momentary disconnections involving bat306301.As a result, the reported event was confirmed.A power source lubrication procedure was performed on (b)(6) 2018 to mitigate the reported conditions and the battery remains in use.A possible root cause of the loss of power can be attributed to a disconnection of both power sources and/or to an intermittent disconnection on one or both power sources.Battery bat306301 h3: yes h6 fda method code(s): 4114, 4112 h6 fda results code(s): 3213 h6 fda conclusion code(s): 12 medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
This supplemental report (1 of 95 corrections identified) is being submitted as a correction to the event date to reflect the date of servicing in the initial report.The initially reported aware date, reporting timelines, reported event and investigation details remain unchanged.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key7756765
MDR Text Key116906439
Report Number3007042319-2018-03510
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00888707000420
UDI-Public00888707000420
Combination Product (y/n)N
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 06/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/30/2017
Device Model Number1420
Device Catalogue Number1420
Was Device Available for Evaluation? No
Date Manufacturer Received06/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1903-2018
Patient Sequence Number1
Treatment
1103 VAD.
Patient Age20 YR
-
-