• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. E-POLY 36MM +3 MAXROM LNR SZ23; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. E-POLY 36MM +3 MAXROM LNR SZ23; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Fracture (1260); Detachment of Device or Device Component (2907)
Patient Problem Pain (1994)
Event Date 07/21/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).The 11-104052 ¿ mallory/head actabular shell ¿ 418660.105423 ¿ acetabular lock ring - 332340.Therapy date: (b)(6) 2015.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the location of the product is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 -018 -06455.0001825034 -2018 -06467.
 
Event Description
It was reported patient underwent left hip revision approximately 6 years post implantation due to pain.It was found that the liner was damaged causing disassociation.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
Reported event was confirmed by review of medical records.Review of op notes confirm that patient was seen for increasing pain, and decreasing function.Catastrophic failure noted on x-ray.Femoral head appeared eccentric and signs of failure in the locking ring apparent.During the revision, the head was severely burnished and was removed.The stem was well-fixed, it was removed; a new cup was implanted with appropriate anteversion abduction angle.No further complications noted.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E-POLY 36MM +3 MAXROM LNR SZ23
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key7758414
MDR Text Key116294792
Report Number0001825034-2018-06461
Device Sequence Number1
Product Code MAY
Combination Product (y/n)N
PMA/PMN Number
PK090103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,study
Type of Report Initial,Followup
Report Date 09/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2019
Device Model NumberN/A
Device Catalogue NumberEP-108223
Device Lot Number205980
Was Device Available for Evaluation? No
Date Manufacturer Received08/21/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age58 YR
Patient Weight72
-
-