Model Number FUSION
Device Problem Poor Quality Image (1408)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/12/2018
Event Type  malfunction  
Manufacturer Narrative
Missing device manufacture date.No parts have been returned to the manufacturer for analysis.
 
Event Description
Medtronic received information regarding a navigation system being used during a functional endoscopic sinus surgery (fess) procedure.It was reported that when the surgeon loaded a patient exam via disc, the images were coming over dark and he was unable to adjust contrast.Issue occurred pre-operatively with less than one hour delay to surgical time.There was no reported impact to patient outcome.
 
Manufacturer Narrative
Additional information: device manufacture date provided.Correction: troubleshooting performed by the surgeon with assistance from technical services (ts) was able to restore functionality through adjustment of the pixel offset settings on the navigation system.A software investigation analysis was initiated to determine the probable cause of the issue through review of the reported issue.Analysis found that the adjustment of the pixel offset settings restored functionality, indicating the exam used led to the reported issue.Analysis found that the software functioned as designed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7760271
MDR Text Key116386639
Report Number1723170-2018-03880
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994380005
UDI-Public00613994380005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/24/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age28 YR
Patient Weight52
-
-