HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number 1650DE
Device Problems Battery Problem (2885); Charging Problem (2892); Power Problem (3010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/18/2018
Event Type  malfunction  
Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Other devices involved in this event: heartware ventricular assist system ¿ battery / (b)(4)/ model #: 1650de / expiration date: 2016-11-30, udi #:(b)(4), mfg date: 2015-11-30, (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient was experiencing power switching while at home.The site further noted that the charge on the batteries, at the time of the event, was greater than or equal to twenty-five percent.The batteries have been exchanged.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Corrected information: medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Manufacturer Narrative
Product event summary: two batteries were not returned for evaluation.Log file analysis revealed that the controller in use during the reported event contained a feature that records whether a power source experienced a communication error or a disconnection within each 15-minute interval.Analysis of the data log file revealed several premature power switching events that were due to momentary disconnections involving (b)(4).As a result, the reported event was confirmed.There is no evidence that the lubrication servicing was performed on the reported devices.Based on the available information, the most likely root cause of the reported event can be attributed to momentary disconnections between the controller and battery.An internal investigation has been opened to evaluate the momentary disconnections and implement corrective actions as required.Additional products: battery, (b)(4).D10: no h3: yes h6 fda method code(s): 4112, 4114 h6 fda results code(s): 213 h6 fda conclusion code(s): 67 medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key7761470
MDR Text Key116707948
Report Number3007042319-2018-03552
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00888707000369
UDI-Public00888707000369
Combination Product (y/n)N
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 03/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/28/2019
Device Model Number1650DE
Device Catalogue Number1650DE
Was Device Available for Evaluation? No
Date Manufacturer Received03/04/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1903-2018
Patient Sequence Number1
Treatment
MCS UNKNOWN VAD
Patient Age69 YR
-
-