Model Number 1104
Device Problems Charging Problem (2892); Pumping Problem (3016)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/28/2017
Event Type  malfunction  
Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.This event was assessed and is being reported as part of a retrospective review of log file data.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient had several low flow alarms.Log file review confirms low flow alarms.The ventricular assist device (vad) remains implanted.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
The pump (b)(4) was not returned for evaluation.The reported "low flow" event was confirmed via log file analysis which revealed a decrease power consumption beginning on (b)(6) 2017 and 17 low flow alarms logged since (b)(6) 2017.There is no evidence to suggest that a device malfunction caused or contributed to the reported event.Based on the risk documentation, possible causes of the reported low flow event may be attributed to multiple factors including but not limited to thrombus at the inflow cannula, constriction at the outflow graft, poor vad filling.The device remains implanted in the patient and is thus not available for return to the manufacturer.With a review of the available information there is no evidence to indicate any device malfunctions or performance issues that would impact the reported events.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM-PUMP
Type of Device
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key7763459
MDR Text Key116865895
Report Number3007042319-2018-03569
Device Sequence Number1
Product Code DSQ
Combination Product (y/n)N
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/31/2018
Device Model Number1104
Device Catalogue Number1104
Was Device Available for Evaluation? No
Date Manufacturer Received08/08/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-