The investigation determined that a lower than expected vitros na+ result was obtained when performing a patient sample correlation on a vitros 4600 chemistry system.The likely cause of lower than expected vitros na+ result is a suboptimal calibration.However, the cause of the suboptimal calibration could not be definitively determined.A diagnostic within run vitros na+ precision test was not performed to evaluate the performance of the vitros 4600 system, therefore, the vitros 4600 system could not be entirely ruled out as a contributor of the event.As vitros na+ lot 4219-0991-0597 had not been put into use, historical vitros na+ quality control results were not available to evaluate the performance of vitros na+ lot 4219-0991-0597.Therefore an issue with the reagent could not be entirely ruled out as a contributor of the event.The lower than expected vitros na+ result was likely caused by a suboptimal calibration event on (b)(6) 2018 likely due to a calibrator handling issue.Acceptable performance was observed using freshly prepared calibrators without altering the condition of the vitros 4600 system.This would indicate that both vitros na+ lot 4219-0991-0597 and the vitros 4600 system did not contribute to the suboptimal calibration.
|