ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP RF CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3262
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Bruise/Contusion (1754); Chest Pain (1776); Unspecified Infection (1930); Swelling (2091)
Event Date 06/19/2018
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that during an office visit on (b)(6) 2018, the patient complained of chest pain and discomfort, racing/skipping heart beats, swelling of hands and feet, cough, cold and heat intolerance, indigestion bruising easily, ringing of the ear difficult swallowing.The right atrial lead exhibited decreased atrial sensing and high capture threshold.The lead was confirmed to be dislodged under fluoroscopy and chest x-ray.The lead was repositioned on (b)(6) 2018.The pacemaker pocket was also revised to a more medial position.During the pocket revision, a yellowish serosanguineous fluid was noted in the pocket and a wound culture was sent out to the laboratory for analysis.The pocket was irrigated with antibiotic solution and the pacemaker was reused.Patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP RF CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7769565
MDR Text Key116643625
Report Number2017865-2018-11554
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734509091
UDI-Public05414734509091
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2019
Device Model NumberPM3262
Device Lot NumberA000045018
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/10/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/02/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
2088TC, (B)(4)
Patient Outcome(s) Required Intervention;
Patient Age72 YR
Patient Weight83
-
-