• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN ARGYLE; TROCAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN ARGYLE; TROCAR Back to Search Results
Model Number 8888565036
Device Problems Leak/Splash (1354); Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/10/2018
Event Type  malfunction  
Manufacturer Narrative
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
 
Event Description
Customer reports the doctor noticed a leak after insertion and upon inspection, there was a hairline tear in the tubing.There was no harm to the patient.
 
Manufacturer Narrative
An investigation was performed for the reported customer complaint: ¿customer reports, the doctor noticed a leak after insertion and upon inspection, there was a hairline tear in the tubing.There was no harm to the patient.¿ a review of the device history record (dhr) for lot no.172090154 indicated the product was released meeting all quality standards.All dhrs are reviewed for accuracy prior to product release.In-process procedures are also in place to prevent nonconforming product in the manufacturing process.This ensures components and finished products meet all quality inspection standards.These controls include, but are not limited to: material verification/certification processes, dimensional specifications, statistical samplings, periodic audits, process inspections, machine maintenance/operation and personnel training and certification.No product/sample was provided for evaluation.No additional information, pictures or videos were received.Therefore, a comprehensive investigation was unable to be conducted.The reported customer complaint could not be confirmed.A root cause could not be determined.This complaint will be used for trending purposes.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARGYLE
Type of Device
TROCAR
Manufacturer (Section D)
COVIDIEN
boulevard insurgentes 19030
tijuana 22225
MX  22225
Manufacturer (Section G)
COVIDIEN
boulevard insurgentes 19030
tijuana 22225
MX   22225
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7770160
MDR Text Key116887041
Report Number1282497-2018-00671
Device Sequence Number1
Product Code DRC
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8888565036
Device Catalogue Number8888565036
Device Lot Number172090154
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/17/2018
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-