SIEMENS HEALTHCARE DIAGNOSTICS RAPIDPOINT 500 BLOOD GAS ANALYZER; RP 500 Back to Search Results
Catalog Number 10492730
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/27/2018
Event Type  malfunction  
Manufacturer Narrative
The customer stated that repeat testing was performed to confirm correct results and a corrected report was issued.Siemens service reviewed the system event log for error messages and "sodium interference" messages were received on the day before and the day of this event.The requested data logs have been received for investigation.The measurement cartridge has been discarded.The customer stated that they installed a new cartridge, the aqc were in the expected ranges and they are operational.The cause of this event is unknown.
 
Event Description
The customer reported discrepant high sodium results on the rp 500.There is no report of injury due to this event.
 
Manufacturer Narrative
Siemens reviewed the data provided by the customer.The na+ sensor was severely impacted by a quartenary ammonium compound such as benzylkonium.A message was displayed in the event log indicating sodium interference was detected by the analyzer.Per the rp500 operators manual, these substances are known interferences to the rp500 na sensor and should be avoided.The customer is taking a cleaner containing benzyl ammonium chloride out of use.The system is operational.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500 BLOOD GAS ANALYZER
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key7770167
MDR Text Key116920982
Report Number3002637618-2018-00095
Device Sequence Number1
Product Code CHL
UDI-Device Identifier00630414589169
UDI-Public00630414589169
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122539
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10492730
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/23/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-