Model Number 9770501
Device Problem Signal Artifact/Noise (1036)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer, at this time, for evaluation.A history review of serial number (b)(4) found a similar complaint reported on work order number (b)(4).The root cause was found to be an internal failure of the probe.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer, at this time, for evaluation.
 
Event Description
Biomed reported on behalf of users that site rite 8 ultrasound system with pinpoint gt has rain artifact in the image.Biomed is unable to duplicate the issue in his office.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample(if available), applicable fmea documents, labeling and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue site rite 8 ultrasound system has rain artifact in the image was unconfirmed.During inspection the scanner was evaluated with the model 550 phantom and the image was found to be normal for a sr8.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested and returned to the customer.
 
Event Description
Biomed reported on behalf of users that site rite 8 ultrasound system with pinpoint gt has rain artifact in the image.Biomed is unable to duplicate the issue in his office.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE 8 ULTRASOUND SYSTEM W/PINPOINT GT
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7770936
MDR Text Key116902689
Report Number3006260740-2018-02036
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741098338
UDI-Public(01)00801741098338
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770501
Device Catalogue Number9770501
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2018
Event Location Hospital
Date Manufacturer Received08/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-