BARD ACCESS SYSTEMS SITE~RITE 8 ULTRASOUND SYSTEM W/PINPOINT GT; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
|
Back to Search Results |
|
Model Number 9770501 |
Device Problem
Signal Artifact/Noise (1036)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been returned to the manufacturer, at this time, for evaluation.A history review of serial number (b)(4) found a similar complaint reported on work order number (b)(4).The root cause was found to be an internal failure of the probe.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer, at this time, for evaluation.
|
|
Event Description
|
Biomed reported on behalf of users that site rite 8 ultrasound system with pinpoint gt has rain artifact in the image.Biomed is unable to duplicate the issue in his office.
|
|
Manufacturer Narrative
|
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample(if available), applicable fmea documents, labeling and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue site rite 8 ultrasound system has rain artifact in the image was unconfirmed.During inspection the scanner was evaluated with the model 550 phantom and the image was found to be normal for a sr8.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested and returned to the customer.
|
|
Event Description
|
Biomed reported on behalf of users that site rite 8 ultrasound system with pinpoint gt has rain artifact in the image.Biomed is unable to duplicate the issue in his office.
|
|