Model Number H7493933515070
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/12/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: returned product consisted of a 7f guidezilla ii guide extension catheter.The device was bloody.The hypotube, collar, distal shaft and tip was microscopically and tactile inspected.Inspection revealed a complete separation at the collar with the separation consistent with torsion force, collar damage (stretched/twisted), and tip damage (misshapen/notched).Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
It was reported that the catheter came apart.The target lesion was located in the superficial femoral artery.A guidezilla¿ ii guide extension catheter was selected for use.During procedure, it was noted that the guidezilla ii came apart inside the patient's body.The device was retrieved and the procedure was completed with a drug coated balloon.No patient complications were reported and patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDEZILLA¿ II GUIDE EXTENSION CATHETER
Type of Device
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7773059
MDR Text Key116867545
Report Number2134265-2018-07039
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163314
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/16/2019
Device Model NumberH7493933515070
Device Lot Number0021617170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/16/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2018
Is the Device Single Use? Yes