The device was not returned to olympus for evaluation.The cause of the reported complaint cannot be confirmed.As a precaution, the device instruction manual recommends, ¿always have a spare instrument available in case the primary instrument malfunctions.¿ pre-procedure inspection instructions in the manual also state, ¿make sure that the adapter biopsy valve, the needle adjuster, the stopper, the connector section and the connecting-slider are not damaged.¿ the pre-procedure inspection also includes operational verification of the needle adjuster and resulting needle extension length.The instruction manual also states that during the procedure, ¿confirm that the needle slider is pulled until it clicks and is locked by the needle adjuster before inserting the instrument into the endoscope.Otherwise, the needle¿s distal end may pierce the endoscope¿s instrument channel or the needle¿s distal end may extend from the distal end of the endoscope abruptly.This could cause patient injury¿.Before insertion, ¿confirm that the needle adjuster is in the proximal position and that the needle adjuster lever is securely tightened.¿ and before taking a sample, ¿before pushing the needle slider, confirm that the needle adjuster is fixed firmly.Be sure to firmly fix the needle adjuster.Otherwise, it could cause excess extension of the needle from the distal end of the sheath.¿.
|
Olympus was informed that during an ebus procedure, the needle adjuster malfunctioned and the needle went into tissue 1 mm deeper than intended.There was no reported patient injury, additional bleeding, or device fragment fallen into the patient.There had been no reported difficulties operating the device or inserting it into the endoscope beforehand.The needle was removed from the patient by withdrawal through the endoscope.The intended procedure was completed.
|