Model Number 102956
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/01/2013
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.The device returned for analysis.The complaint investigation determined the reported difficulty was the result of a design related wear issue.After review of this event and similar incidents, abbott has decided to initiate a voluntary field action for centrimag.Abbott performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.
 
Event Description
It was reported that the centrimag displayed an error message that the set pump speed was not reached.The unit was exchanged out to resolve the issue.This event was previously not reported due to backup system being used and no patient was harmed.Abbott made the decision on (b)(6) 2018 to initiate a voluntary recall, therefore this event is being upgraded to reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG MOTOR, US
Type of Device
CENTRIMAG MOTOR
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key7775221
MDR Text Key116863547
Report Number2916596-2018-03461
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 08/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number102956
Device Catalogue Number102956
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2013
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/13/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-