• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 05/08/2018
Event Type  Malfunction  
Manufacturer Narrative

There was no patient involvement. Initial contact telephone number: (b)(6). The heater-cooler 16-02-80 is not distributed in the usa. However, it is similar to the heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601). Livanova (b)(4) manufactures the heater-cooler system 3t. The event occurred in (b)(6). A livanova field service representative was dispatched to the facility to investigate. The service representative was able to confirm the reported issue and identified the root cause to be a faulty motor. The stirrer motor, processor board and distribution board were replaced to resolve the issue. Subsequent functional testing was completed without further issues and the unit was returned to service. A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue. If any additional information pertinent to the reported issue is received, it will be provided in a supplemental report. Device not available for return.

 
Event Description

Livanova (b)(4) received a report that the stirrer motor of a heater-cooler system 3t was stuck and could not be turned during priming. There was no patient involvement.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7782095
MDR Text Key117192628
Report Number9611109-2018-01163
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation
Remedial Action Other
Type of Report Initial
Report Date 08/14/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received08/14/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number16-02-80
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received05/08/2018
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured05/02/2016
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-