• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CONFIRM; IMPLANTABLE CARDIAC MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CONFIRM; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number DM3500
Device Problems Migration or Expulsion of Device (1395); Under-Sensing (1661)
Patient Problem Twiddlers Syndrome (2114)
Event Date 08/07/2018
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented in clinic.Upon interrogation, the implantable cardiac monitor exhibited undersensing due to loss of electrode contact in the pocket.When checking the device, it appeared to be perpendicular to the skin.It was noted that the patient had twiddled with the device.After touching it the device returned to its correct orientation and sensing returned to normal.Patient was in good condition throughout.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7782144
MDR Text Key117209405
Report Number2017865-2018-11789
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067027320
UDI-Public05415067027320
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/22/2019
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberS000057303
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/28/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-