DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22495
Device Problem Defective Alarm (1014)
Patient Problems Hypoglycemia (1912); Loss of consciousness (2418)
Event Date 07/19/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Diabetes mellitus is a known cause of hypoglycemia and loss of consciousness.
 
Event Description
Dexcom was made aware on (b)(4) 2018, that on (b)(6) 2018, an intermittent audio output and an adverse event was reported.Date of issue is an approximation.The patient stated they experienced a low blood glucose event in which they became comatose.The patient indicated that the receiver did not provide alerts.Emergency medical technician¿s (emts) were called and transported the patient to the hospital via ambulance.During the ambulance ride, the patient was treated with intravenous (iv) therapy of glucose and they recovered consciousness in the ambulance.The patient was admitted to the emergency room (er) and released the same day.At the time of contact, the patient had recovered from the event.No additional patient or event information is available.The complaint device was returned for evaluation.An exterior physical visual inspection was performed, and no defects were found.The receiver was charged and rebooted successfully.A review of the downloaded receiver log did not find any errors related to the customer complaint.Functional testing was performed, the test passed.The receiver case was opened for further evaluation.An interior inspection was performed, and the inspection passed.A try it test was performed, and the test passed.The receiver was tested on the global receiver communication tool and the test passed.The resistance of the speaker was measured, and the resistance was within specification.The reported intermittent audio output could not be confirmed.A probable cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key7782941
MDR Text Key117131729
Report Number3004753838-2018-092058
Device Sequence Number1
Product Code MDS
UDI-Device Identifier30386270000157
UDI-Public30386270000157
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 07/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22495
Device Catalogue NumberSTR-DR-001
Device Lot Number5232899
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/01/2018
Was the Report Sent to FDA? No
Date Manufacturer Received07/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/26/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age80 YR
Patient Weight77
-
-