Catalog Number 9735665
Device Problems No Display/Image (1183); Electrical /Electronic Property Problem (1198)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/20/2018
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment. Testing revealed that the monitor was not locked and in an asleep state. By touching the controls on the bottom right the monitor was functional again. The monitor was locked on the proper input. The system then passed the system checkout and was found to be fully functional.
 
Event Description
Medtronic received information regarding a navigation system being used during a functional endoscopic sinus surgery (fess). It was reported that after registering and proceeding to navigation the main cart monitor went black. No further information could be provided or troubleshooting performed during case. Issue occurred intra/peri-operatively with less than one hour delay to surgical time. There was no reported impact to patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION S8 PREMIUM SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
maxwell roth
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7783505
MDR Text Key117153919
Report Number1723170-2018-04054
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public00643169722187
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial
Report Date 08/14/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number9735665
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/20/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/26/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-