• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE MAHWAH CARDIOSAVE SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE MAHWAH CARDIOSAVE SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problems Failure to Power Up (1476); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/22/2018
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this iabp unit is not required to be reviewed per (b)(4) since the device manufacture date is greater than one year from the event date. A getinge field service engineer (fse) evaluated this unit and found fault code #139 in diagnostics logs. This is a communication error between the power management board and executive processor board. After restart of unit this fault did not appear again. The fse completed pm and functional testing and safety check to factory specifications. The iabp passed all functional and safety tests per factory specifications and was returned to customer and cleared for clinical use.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) unit failed during the start-up procedure, it has freezed during the booting. The iabp was leaved over and another unit was used by customer. There was no patient involvement and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7785263
MDR Text Key117430192
Report Number2249723-2018-01391
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/15/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Device Age YR
Event Location No Information
Date Manufacturer Received07/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-