Device Problem Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/24/2018
Event Type  malfunction  
Manufacturer Narrative
A gehc service representative performed a checkout of the equipment and confirmed the reported complaint. The random access memory was cleaned and reset. Patient information could not be obtained due to country privacy laws. The initial reporter is located outside the u. S. , and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported that, during a case, the machine shut down. The anesthesia machine was replaced. There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAVANCE
Type of DeviceANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key7800711
MDR Text Key117705950
Report Number2112667-2018-01645
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K032803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 08/21/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/12/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-