Model Number N/A
Device Problem Power Problem (3010)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) unit was reviewed and no non-conformances related to the reported event were noted.(b)(6).A getinge field service engineer (fse) evaluated the iabp unit and was able to confirm the reported problem.The iabp unit would not turn on and the batteries had been depleted.While troubleshooting the fse noticed the console was not making full contact with the cart.This caused the batteries to deplete and not allowing the unit to power on.The fse reseated the console to the cart and allowed the batteries to charge.The customers asked the fse to replace the batteries.The iabp was then released to the customer for return to clinical service.
 
Event Description
It was reported that prior to use on a patient, the cardiosave intra-aortic balloon pump (iabp) had no a/c power.No adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7802725
MDR Text Key117852304
Report Number2249723-2018-01442
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 08/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/07/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse