• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY ESCORT II EXTRACTION BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY ESCORT II EXTRACTION BALLOON Back to Search Results
Catalog Number EBL-12-200
Device Problem Deflation Problem (1149)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/30/2018
Event Type  malfunction  
Manufacturer Narrative
Common name: biliary catheter for stone removal that may also allow for irrigation and contrast injection.Investigation evaluation: our laboratory evaluation of the product said to be involved could not confirm the report as it was described.A functional test was performed on the balloon.The prepackaged syringe that was supplied with the balloon was attached to the inflation port.When the balloon was filled with air, no leaks were present in the balloon material.Once the balloon was inflated, pressure was released from the syringe and the balloon deflated within a few seconds.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for the reported observation could not be determined because the product said to be involved functioned as intended.A discrepancy or anomaly that could have contributed to the reported observation was not observed during our laboratory analysis of the returned product.The information provided indicated the balloon functioned properly prior to use.Therefore, the balloon was functioning prior to advancement through the endoscope.The instructions for use (ifu) states, "once balloon is endoscopically visualized in duodenum, turn stopcock to open position and deflate balloon." prior to distribution, all escort ii extraction balloon are subjected to a visual and functional test to ensure device integrity.The functional test includes an air inflation test to ensure proper balloon function.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted and this represents an isolated occurrence.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook escort ii extraction balloon (ebl-12-200).The ebl balloon did not deballooning [deflate] in the bile duct after the bile duct stone removal procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESCORT II EXTRACTION BALLOON
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key7812587
MDR Text Key118406235
Report Number1037905-2018-00378
Device Sequence Number1
Product Code GCA
UDI-Device Identifier00827002216037
UDI-Public(01)00827002216037(17)190424(10)W4059220
Combination Product (y/n)N
PMA/PMN Number
K953951
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 07/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/24/2019
Device Catalogue NumberEBL-12-200
Device Lot NumberW4059220
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/24/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age64 YR
Patient Weight76
-
-