LIVANOVA USA, INC. LEAD MODEL 304 Back to Search Results
Model Number 304-20
Device Problems Fracture (1260); Mechanical Problem (1384); Device Contamination with Body Fluid (2317)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/18/2018
Event Type  malfunction  
Event Description
High impedance on the patient's device was detected through a periodic review of the manufacturer's programming history database.No further relevant information has been received to date.No known relevant surgical intervention has occurred to date.
 
Event Description
It was reported that the patient underwent a full revision of generator and lead due to high impedance.The suspect product was discarded.No further relevant information has been received to date.
 
Event Description
The suspect device was received for analysis, but analysis has not been completed on the device to date.
 
Event Description
Product analysis was completed on the returned generator.The generator performed as expected with no performance anomalies identified product analysis was completed on the returned lead.The electrode array wasn't received so an evaluation and resulting commentary cannot be made on that portion of the lead.A lead fracture in the negative coil at the point of an abraded opening in the inner and outer tubing was identified.The negative coil had what appeared to be wear (flat surfaces) in the vicinity of the break location.The lead assembly has dried remnants of what appear to have once been body fluids inside the inner and the outer silicone tubing.No obvious point of entrance was noted other than the identified tubing openings and the cut end of the returned lead portion.A suture was also present constricting the lead, at which point the lead coils were kinked/nicked no further relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 304
Type of Device
LEAD
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key7814165
MDR Text Key118254117
Report Number1644487-2018-01484
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 12/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/29/2016
Device Model Number304-20
Device Lot Number202017
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2018
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received11/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/06/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age18 YR
-
-