Model Number S7 |
Device Problems
Loss of Power (1475); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/01/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.Missing udi.Missing device manufacture date.No parts have been returned to the manufacturer for analysis.
|
|
Event Description
|
Medtronic received information regarding a navigation system.It was reported outside of a procedure that the screen display of the touch screen monitor had a blackout.There was no patient present.
|
|
Manufacturer Narrative
|
Additional information: unique device identification (udi) and device manufacture date provided.Correction: product and related fields updated to proper value.Correction: no 510(k) provided as this device is not released for distribution in the united states.
|
|
Manufacturer Narrative
|
The monitor was returned to the manufacturer for analysis.Analysis found that the monitor powered up with a normal display and touch function was normal.However, after warming up, the display went blank several times and then returned to a normal display.Analysis found that the reported event was related to an electrical issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
|
|