Catalog Number 08H67-01
Device Problem Explosion (4006)
Patient Problem No Patient Involvement (2645)
Event Date 07/31/2018
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A followup report will be submitted when the evaluation is complete.Evaluation in process.
 
Event Description
The account stated the cell-dyn ruby analyzer had an explosion with soot on the top of the analyzer and monitor.The cell-dyn ruby analyzer is off but the monitor is on with a message of serious error code 2077.The power switch was then turned off on the cell-dyn ruby analyzer and the monitor.No patient involved.No injury or exposure was reported.
 
Manufacturer Narrative
The investigation consisted of review of the complaint text, product historical data, product labeling, and consultation with on-market engineering.Review of product historical data for any trends and all customer complaints received for this issue did not identify any adverse trends or abnormal complaint activity.The power supply is not available for return for investigation, since the unit was discarded.The service engineer stated that the power supply unit replaced did not present any residues or traces of explosion, or flames.A review of the product labeling concluded that the issue is sufficiently addressed.No product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN RUBY ANALYZER
Type of Device
AUTOMATED HEMATOLOGY ANALYZER
Manufacturer (Section D)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
MDR Report Key7818297
MDR Text Key118750823
Report Number2919069-2018-00042
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier00380740017170
UDI-Public00380740017170
Combination Product (y/n)N
PMA/PMN Number
K061667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number08H67-01
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/27/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-