It was reported that the patient underwent vns explantation surgery due to lack of efficacy.The explanted products were received by the manufacturer.Generator product analysis was completed.The reported lack of efficacy could not be evaluated in the product analysis, or pa, lab however, the observed depleted battery could have been a contributing factor.It was observed that the supply current pulsing was out of specification during the post burn test.Analysis indicated that during manufacturing, the r35 resistor was not optimally chosen.A lower value resistor would have more suitably centered the currents within the limits.With r35 reselected in the pa lab, the device performed according to functional specification.The 2,868 o resistor was replace with a 2,671 o resistor and test specifications were met.The out of specification supply current pulsing could have potentially been a contributing factor to the end of service condition.However, the results of the battery longevity calculation indicated the eos condition was an expected event.The lead product analysis will not be reported as it was not relevant to the out of specification generator resistor.No additional relevant information has been received to date.
|