ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC; PACEMAKER DATA TRANSMITTER Back to Search Results
Model Number EX1150
Device Problem Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/30/2018
Event Type  malfunction  
Manufacturer Narrative
Final analysis found burn marks on the inner casing of the ac power adapter and burnt components on the printed circuit board of the ac power adapter and the transmitter, which contributed to the field complaint of a power up anomaly.It was concluded that the transmitter was hit by a high current flow through the ac wall power outlet into the ac power adapter and through the telephone line into the transmitter, thus damaging their respective main pcb.
 
Event Description
It was reported that the patient's merlin at home transmitter would not power up after a storm.Reconnecting the prongs to the adapter and trying a different power outlet failed to resolve the issue.The transmitter was replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7825109
MDR Text Key118557407
Report Number2017865-2018-12539
Device Sequence Number1
Product Code DRG
UDI-Device Identifier05414734504799
UDI-Public05414734504799
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial
Report Date 08/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEX1150
Device Catalogue NumberEX1150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2018
Was the Report Sent to FDA? No
Date Manufacturer Received08/16/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/27/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-